Model Number DEX3100S13
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2026
Type of Reportable Event Malfunction
Event or Problem Description
A bipap a40 pro device was returned to a third-party service center for service.There was no serious patient harm or injury that occurred or was reported.During the evaluation at the third-party service center, a ventilator inoperative error code was found in the device's error log stating the difference between the blower pressure sensor reading and the outlet pressure sensor reading is 10 cmh2o or greater for 5 seconds.There is no documentation stated on a root cause and/or any component replacement to resolve the issue.No repairs were performed.The device will be scrapped per the customer's request.
 
Additional Manufacturer Narrative
Dhr review was performed for the complaint device sn.Review confirms that the device met the final acceptance criteria and was released for final distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Bipap A40 Pro
Common Device Name
Ventilator, Continuous, Minimal Ventilatory Support, Facility Use
Manufacturer (Section D)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
+1(724)733-0200
MDR Report Key50945977
Report Number2518422-2026-046646
Device Sequence Number10451977
Product Code MNT
UDI-Device Identifier00606959055186
UDI-Public(01)00606959055186
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K121623
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDEX3100S13
Device Catalogue NumberDEX3100S13
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/24/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1813-2024
Patient Sequence Number1
Patient SexUnknown
-
-