|
It was reported that the driving cap f/insertion handle, multihole drill sleeve, trocar yell, scrdriver-hex 5 flex cann, extract-instr f/tfna helical blade+scr, scrdriver short xl25, scrdriver w/quick-coupl hex 12 short xl2, hand lrg cann w/quick-coupl hex12, connecscr f/tfna helical blade+scr, and scrdriver shaft self-holding l165 f/star were associated with wear-related and deformation issues.The reported failures included stripped or worn threads that did not engage securely, deformed holes that made guide wire passage difficult, a bent trocar that did not fit the cannula properly, dull screwdriver features that did not seat securely or function properly, and a handpiece that no longer clicked into screwdriver shafts.There was no patient involvement.The complaint devices were retained by the customer.
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11.Additional manufacturer narrative:
d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
|