| Model Number |
3501 |
| Medical Device Problem Codes |
Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Read Input Signal (1581)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
09/25/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient had an additional inappropriate shock due to t-wave oversensing via wide intrinsic complexes.The sensing vector was set to the alternate vector.The office testing found poor sensing in all three of the sensing vectors.The doctor elected to program the therapy off and sent the patient home with a life vest.The plan is to explant and replace the system with a transvenous system.There were no additional adverse patient effects.The system remains implanted.
|
| |
|
Search Alerts/Recalls
|