The account reported finding the device had slow humidification rate.Medline quality received 1 unit; visually inspected the sample and it was observed that the filter was due for replacement, the sample turned on and did mist, the digital flowmeter had a reading of 7.8 lpm, and the unit ran at 30 minute interval without issues with temperatures that met product specifications.The reported issue was not confirmed with received sample, and medline was not able to confirm what may have caused the issue.The quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|