|
The account reported check valve backwards and tubing too tight, and that the tubing is knotted together.No stock check was able to be performed as there was no remaining inventory; a photo was received for review, and review of the finished good bill of materials/pack build, finished good manufacturing instructions, and device history record did not reveal any nonconformances related to the reported defect.Based on review of the photo, the reported defect was confirmed as the tubing was wound too tightly; the check valve was placed correctly.Root cause can be attributed to human error, and no further action is required at this time as medline will continue to monitor for additional occurrences and adverse trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|