It was reported that this right atrial (ra) lead was a case of an attempted implant due to terminal pin anomaly on the ra lead.As the ra lead was implanted without anomalies and test via pacing system analyzer (psa), however, when ra lead was placed in the right ventricular (rv) lead port of the header a small blips or non physiologic pulses on the ra channel was noted.The ra lead terminal was removed and placed back on the psa with normal tracing, however, when placed back in the ra port of the device baseline blips still present and not oversensed.Furthermore, the impedance and sensing were normal.It was then confirmed that the anomaly was related to the terminal pin when rv lead was placed in the ra port and there were no blips of noise were seen.This ra lead was replaced with the same model, not implanted and will be returned for analysis.No adverse patient effects were reported.
|