According to the customer, there is ¿leaking from the blue neck leading to the needle¿ causing ¿blood to leak from the device¿; the customer reported there was no patient or user impact and no serious injury, medical intervention, or follow up care.Medline received 1cs, llad luer lock access device in new condition, randomly selected 20 samples for visual and functional testing, and used a 21g medline blood collection set with a venipucture blood board to simulate blood sample collection while observing for fluid leakage.A visual inspection revealed no manufacturing defects, functional testing showed no leaking, and the complaint was not confirmed.The samples worked as intended, and based on the review, medline was unable to confirm the complaint issue or that the product failed to meet specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|