The account reported via email that there was an area that did not inflate and caused sinking where the reporter¿s mom¿s bed sore was; no report of any patient injury or adverse event was received.Medline was provided the sample, pumped the mattress up, and sprayed soapy water on the mattress; no leaking cells were found.The defect of the mattress not fully inflated was unable to be confirmed, and the cause was not confirmed with the received sample.No defect was found, corrective action type was none, and there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|