Product complaint # (b)(4).D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.The following information was requested, but unavailable:
- could you please provide the lot number? unk.H3 investigation summary.The product was returned to ethicon for evaluation.One paper lid, one winding former with a needle-suture piece inserted in a yellow sponge and one suture piece outside them were received for analysis.Product code vcp494g.During the initial examination of the returned sample, no evidence of suture breakage was observed.However, the ends of the suture segments were found to be severed, with the cut consistent with contact from a surgical instrument.Biological residue was present on the returned suture segments.As per the conditions of the returned sample, no conclusion could be reached on the cause of the reported event.The instructions for use contain the following caution: as with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of surgical instruments, such as needle holders and forceps.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.Device history review.The manufacturing records could not be reviewed as the batch/lot number is unknown.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|