| Model Number |
7840 |
| Medical Device Problem Codes |
Failure to Capture (1081); High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that this right atrial (ra) lead exhibited a loss of capture on the presenting electrogram.Additionally, a lead safety switch occurred that resulted in high out of range pace impedance measurements.Technical services (ts) was consulted to review the system and found that the anomalies occurred while in bipolar mode.Ts recommended adjusting to a unipolar pace and bipolar sense which resolved the loss of capture with full sensing available.Ts noted that the health care professional was going to discuss further with the cardiologist and assess the chest x ray that was completed.No adverse patient effects were reported.This ra lead remains in service.
|
| |
|