The account reported finding broken manifold during a power injection.One sample perceptor manifold was received for evaluation; finished good production records, finished good pack build, and trending were reviewed.The visual inspection found the rotating adaptor (ra) to be separated from the manifold body, confirming the reported issue.Manifold specs are 200 psi static, 750 psi dynamic, not intended for use with power injection, and based on the complaint intake reporting of the event provided by the customer, this 200s manifold was being used outside of its intended design specs, with a power injector; there was no confirmed medical intervention or serious injury reported related to this incident.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|