Catalog Number MPH6540
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/26/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this one is out of range >50 i.E.57 & 58.
 
Additional Manufacturer Narrative
The customer reported ¿3 meters 3 different problems¿ and that this one was ¿out of range >50 i.E.57 & 58¿; there was no reported death or serious injury associated with this event.A sample/photo was not provided for this complaint, and investigation included review of recent complaint trends/scar activity for the last three trend reports; the vendor was notified.The investigation results were not confirmed, and the complaint was not confirmed because the issue can only be confirmed by further investigating the sample or a provided photo.The cause was documented as sample required to determine cause, and no trends were identified for the specified material number and its corresponding defect type.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
METER, GLUCOSE, HARMONY,PROF.USE ONLY
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
joseph weiske
three lakes drive
northfield, IL 60093--275
MDR Report Key50946341
Report Number1417592-2026-54232
Device Sequence Number21444246
Product Code LFR
UDI-Device Identifier10888277675674
UDI-Public10888277675674
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K161299
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMPH6540
Device Lot NumberNH039213
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/26/2025
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-