• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Black Diamond Video, Inc. Harmony iQ; Harmony iQ Integration System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Black Diamond Video, Inc. Harmony iQ; Harmony iQ Integration System Back to Search Results
Model Number Harmony iQ 2800
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported that the monitor connected to their harmony iq integration system was frozen on a video display.This event did not occur during a patient procedure.No report of injury.
 
Additional Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the unit and found that the iq cpu stack required a reboot.To resolve the issue, the technician rebooted the iq cpu stack, tested the unit, confirmed it to be operating according to specifications, and returned it to service.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Harmony iQ
Common Device Name
Harmony iQ Integration System
Manufacturer (Section D)
Black Diamond Video, Inc.
503 canal blvd
richmond CA 94804
Manufacturer (Section G)
Black Diamond Video, Inc.
503 canal blvd
richmond CA 94804
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
+1(440)259-7401
MDR Report Key50946345
Report Number3008776287-2026-00698
Device Sequence Number10446633
Product Code FTA
UDI-Device Identifier00724995199135
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHarmony iQ 2800
Device Catalogue NumberRLQ2000SLC
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-