The account reported a leaking item in the epidural tray and stated it has been happening quite a bit.A photo and a sample were received; the pack arrived open, there was a large stain on the csr wrap, and upon unwrapping the pack, the chloraprep applicator was seen as the reason for the stain; the applicator was still sealed but had been prematurely activated, and other components had some of the chloraprep stain on them.Finished good production records and pack build history were reviewed with no line down situations, run without approvals, substitutions, deviations, non-conformance reports, or pack build issues that might have had an impact; based on the sample evaluation, the reported problem was confirmed.The placement of the chloraprep in the pack was reviewed and its location may have contributed to the pre-activation, components may also become damaged during transit or storage, and there is not enough information to provide a true root cause; the quality team will monitor this issue for trending purposes and production and qc managers have been notified.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|