Per customer call, the wrist blood pressure unit was giving inaccurate readings.Medline received one automatic digital blood pressure monitor with wrist cuff; upon visual inspection there were no noticeable defects or cosmetic damage, and functional testing was performed by simulating blood pressure at 120/80 @ 80bpm and 154/100 @ 80bpm, with values noted as 120/80 @ 80bpm and 153/100 @ 80bpm.These are within acceptable ranges, and the complaint was not confirmed.The cause was documented as unconfirmed defect - not confirmed with received sample, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|