The customer reported receiving a medline handheld aneroid that is leaking air.Medline received 1ea handheld aneroid, visually inspected the sample and found no obvious signs of defects, and attempted to inflate the cuff to 300mmhg; however, the cuff did not inflate.The complaint was confirmed, and the bladder has a leak that will not allow the cuff to inflate.The identified cause was "comp./part defective, broke," and the division will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|