A lead extraction procedure commenced to remove a right ventricular (rv) lead due to fracture.A right atrial (ra) lead was present within the patient, but was not targeted for extraction.A spectranetics lld ez lead locking devices (lld ez) was inserted into the lead to provide traction.During extraction, utilizing multiple spectranetics devices (glidelight laser sheath, visisheath dilator sheath, tightrail sub-c rotating dilator sheath, and tightrail rotating dilator sheath), the lead, along with the lld broke in the subclavian.Switching to a merit medical needle's eye snare, attempts were made to snare from the groin; however, the snare broke and a abbott agilis catheter was then used.After several attempts to dislodge the lead, a pericardial effusion was noted in the rv apex, and rescue efforts began, including pericardiocentesis and sternotomy.An rv perforation was discovered and repaired, and the rv lead was removed.The patient was transferred to the icu in critical condition, and made no improvement in the following days.Care was withdrawn and the patient did not survive.This report captures the lld that broke, and perforation occurred, requiring intervention, but resulting in death.
|
|
B6/b7) patient information regarding relevant tests/laboratory data or medical history - not available from facility.H3) the lld was discarded, thus no product investigation was performed.H6) per ifu, perforation and death are listed as potential adverse events with use of the lld device.Additional health effect impact code: f15 recognised device or procedural complication.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|