• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; WHLCHAIR, K3, 20", SB DDLA, SA FOOT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; WHLCHAIR, K3, 20", SB DDLA, SA FOOT Back to Search Results
Catalog Number K3206N24S
Medical Device Problem Codes Detachment of Device or Device Component (2907); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "tires are worn off, right wheel is coming off, bent, not safe to use.".
 
Additional Manufacturer Narrative
The customer reported the tires are worn off, the right wheel is coming off, bent, and not safe to use.The complaint could not be confirmed as there were no lot/serial numbers, samples, pictures or videos received.Without a sample received, it is difficult to determine a true root cause for the reported issue.Should the sample/detailed picture become available, the complaint will be reopened and investigated accordingly.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
WHLCHAIR, K3, 20", SB DDLA, SA FOOT
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
diana arizmendi
three lakes drive
northfield, IL 60093--275
MDR Report Key50946505
Report Number1417592-2026-54274
Device Sequence Number10451608
Product Code IOR
UDI-Device Identifier00193489087164
UDI-Public00193489087164
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K990463
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberK3206N24S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2023
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexFemale
-
-