| Catalog Number |
07682913001 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457); Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter received a questionable elecsys total psa assay result for one patient sample tested on the cobas e 801 analytical unit.On 03-sep-2026:
initial analysis yielded 21.3 ng/ml.On 07-sep-2026:
repeat analysis yielded 7.26 ng/ml.Repeat analysis using a different sample tube yielded 6.94 and 7.11 ng/ml.
|
| |
|
Additional Manufacturer Narrative
|
|
The elecsys total psa reagent lot number is 88732700.The expiration date was not provided.The field service engineer inspected the analyzer and replaced the measuring cell, sipper nozzles, and pinch valves, and performed a successful instrument check.The investigation determined the event was due to a component failure.The specific cause of the event could not be determined.The customer has not reported any other issues after the service.The investigation determined that the service actions resolved the issue.
|
| |
|