It was reported that during qc of a new bottle of strips, "the low control is reading high and high reading low," and it "was done multiple times by 2 different staff members." medline received the harmony blood glucose meter in used condition; the glucose control solutions and test strips used by the account were not provided, and medline tested the sample using harmony glucose control solutions and harmony blood glucose test strips, with level 1 and level 3 results of 47 mg/dl and 211 mg/dl, respectively, within test ranges.The complaint was not confirmed, and the cause code was not confirmed with received sample.Medline saw no issues during testing, the sample turned on as intended, readings were within the control solutions range, and the sample worked as intended.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|