The customer/reporter described that hemo broke during use and that the hemostat broke in half when inserting a chest tube.Finished good production records were reviewed and there were no run without approvals, deviations or non-conformance reports that might have had an impact on this issue; finished good pack build was reviewed and there are no issues with the build that likely contributed to the reported issue.A sample was received, and the hemostat received the left handle was broken into two pieces confirming the reported issue.Further investigation reveals that the root cause is attributed to a vendor product issue related to the component manufacturing process; however, given that this is a multi-sourced component, the vendor could not be definitively identified, and the supplier quality team will continue to monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|