Model Number PT-000409
Medical Device Problem Code Delayed Program or Algorithm Execution (4034)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 09/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when a bolus was commanded on the omnipod dash controller/ personal diabetes manager (pdm) it would not deliver until hours later, resulting in the patient receiving a sudden rush of insulin.The patient¿s blood glucose levels were approximately 20 mmol/l (360 mg/dl) when they administered the bolus.When the bolus didn¿t initially deliver, the patient administered manual injections, but this had caused ¿dangerous lows¿ when the bolus delivered later.No medical assistance was required.
 
Additional Manufacturer Narrative
The device has not been returned/received to date.If the device is received, a supplemental report will be submitted with the investigation results.We are unable to determine if any product condition could have contributed to the reported event.Lot release records were reviewed and the product lot met all acceptance criteria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Omnipod 5 Controller
Common Device Name
Alternate Controller Enabled Insulin Infusion Pump
Manufacturer (Section D)
Insulet Corporation
100 nagog park
acton MA 01720
Manufacturer (Section G)
Insulet Corporation
100 nagog park
acton MA 01720
Manufacturer Contact
derek heyl
100 nagog park
acton, MA 01720
+1(978)600-7000
MDR Report Key50946569
Report Number3014585508-2026-72386
Device Sequence Number18749162
Product Code QFG
UDI-Device Identifier10385083000022
UDI-Public(01)10385083000022(10)L000529
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K203768
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberPT-000409
Device Catalogue NumberPDM-H001-G-XX
Device Lot NumberL000529
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/30/2026
Date Device Manufactured02/06/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexFemale
-
-