This vmsr submission is part of a retrospective data remediation review.The included event represents one malfunction event that medline became aware of during q3 2025.Associated lot numbers: 65122100002 and 65124080001.No serial numbers were available.No device identifier (di) information was available.One malfunction event involving two devices is summarized in this report.Two devices were returned to the manufacturer and evaluated.The devices' labeled single-use status was not available.No devices were identified as reprocessed or reused.The account reported receiving two medline hand held aneroids, catalog number mds9380, that were leaking air.The complaint was reported after the customer indicated that two aneroids had been returned because they were leaking air and replacement devices were requested.Medline quality received and evaluated two devices.Visual inspection identified no obvious external defects.Functional testing was performed by attempting to inflate the cuffs to 300 mmhg; however, neither cuff could be inflated.Evaluation confirmed that both devices contained bladder leaks that prevented cuff inflation, confirming the reported malfunction.The affected sku and lot numbers were found to fall within scar-00444, which had been previously initiated to address leakage issues.The complaint was confirmed and the investigation determined that the failure was associated with the known leakage condition addressed under scar-00444.Medline continues to work with the vendor to implement corrective actions.The division will continue to monitor complaint trends.Trending reviews were completed under qual-144184, qual-140946, and qual-138405.Trends associated with the specified material number and corresponding defect type were identified.A replacement was issued under order (b)(6).There were no reports of death, serious injury, medical intervention, or follow-up medical care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable malfunction event.If additional information becomes available, this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|