The customer complaint reported the enfit stopcock was defective, with a crack noticed where the luer lock is and another port ended up breaking.A sample of enfit9000ns was received from the customer; the sample was in a very used condition with the print on the valve plate slightly faded and dried formula in the connectors.Based on the sample provided, the reported issue was confirmed, and the breaks were located on both male enfit connectors with notable damage to one of the enfit caps that get inserted into the connector.The cause was documented as cannot determine cause with received sample, and the divisional quality representative will continue to monitor this issue for tracking and trending purposes during quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|