Catalog Number MDS1001
Medical Device Problem Codes Computer Software Problem (1112); Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/25/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was "returned by the customer as not functioning"; "the motherboard is not working just as the one before"; and "we saw an error message on the screen which is a \" 3+ \".".
 
Additional Manufacturer Narrative
The customer explained that the unit was returned by a customer as not functioning, the motherboard is not working per the customer technician, and they received an error message on the screen "3+".Samples were not received for investigation; however, photos were provided that showed the display issues reported, and the vendor stated, "after checking the process of our company, we found no problems related to the process leading to complaints, but our company will record and observe the trend." the complaint was confirmed; the issue shown via investigation was caused by a functional malfunction preventing the monitor to work correctly, and the cause was documented as "cannot determine cause w/received sample." material trending via sap showed similar complaints during the last 12 months, and division and vendor have been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBD-MONITOR, BP, AUTO, DIGITAL, ADULT
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
rylee rosenberg
three lakes drive
northfield, IL 60093--275
MDR Report Key50946713
Report Number1417592-2026-54319
Device Sequence Number9534655
Product Code DXN
UDI-Device Identifier00888277629656
UDI-Public00888277629656
Combination Product (Y/N)N
Initial Reporter CountryPM
PMA/510(K) Number
K193627
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS1001
Device Lot Number230580005
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2024
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-