During clipping in the sigmoid post polyp removal, the physician used a cook instinct plus endoscopic clipping device.It was reported that the clip closed perfectly but once "[the user]" crunched, the clip did not deploy, (heard the deployment) but the clip was still on the catheter.Another clip was used and deployed perfect as it should.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
|
Continued:
e1.Phone: (b)(6).Investigation evaluation:
a product evaluation was performed only by the photos provided in response to this report because the product said to be involved was not provided to cook for evaluation.Per the photos provided we cannot complete a full evaluation.Without the product or substantial evidence to contradict the complaint, it is considered confirmed based solely on statements and photos describing the event.The first photo shows the device coiled inside the device pouch, the lot number on the label matches that in the report.The second photo shows the distal end of the device; the clip is partially opened.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion:
our evaluation of the photos provided could not confirm the report.Additionally, we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use states, "note: if separation of the clip is not immediate, gently move the catheter back and forth or use other endoscopic maneuvers to separate the catheter from the clip."
prior to distribution, all instinct plus endoscopic clipping devices are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action:
based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
|