The account reported finding rusting blades and stated blades are rusted in the solution in compartment, while other tools in compartment made from carbon-steel are not rusting only this item.A photo and a sample were received for investigation; two boxes of unused items were received, visual inspection identified no issues, and the items appeared brand new and remained in their original packaging with no signs of damage or tampering.Pictures were received, and visual inspection of the images revealed that the customer was subjecting disposable blades designed for single use to a cleaning and sterilization process alongside reusable handles; based on the photos provided by the customer, the complaint of rusting blades is visible therefore confirmed.Finished good production records were reviewed and did not find any nonconformances, the updated cause code was adjusted to dc: traced to user with misuse/mishandling, and the division will continue to monitor the issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|