| Model Number |
LifeVest WCD 4000 System |
| Medical Device Problem Code |
Charging Problem (2892)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor reported that a battery charger was resetting.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of battery charger has been completed.The reported problem (resets) was isolated to the bga failure of pld component and pxa component on the c/a board.The root cause for the component failures could not be positively identified.There was no adverse event that resulted from the damaged charger.
|
| |
|