The customer reported having "3 hudson rci units that are in need of repair" and stated, "as far as i am aware, all 3 are just no longer heating." medline quality visually inspected the sample and found the unit in good condition; the breathing circuit, airway temperature probes, heated wire cables, concha column and concha water reservoir were connected for further testing, and the unit was plugged into a power source with a confirmed red/yellow indicator flash.The unit was powered on but immediately powered itself off; as a result, medline cannot confirm the reported complaint.The root cause cannot be identified, and the division will monitor the complaint during quarterly trending and notify the manufacturing site of the issue on a quarterly basis.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|