The patient had primary right shoulder surgery on (b)(6) 2022.On (b)(6) 2023, the patient came in reporting pain due to a malpositioned baseplate and the cause of the malpositioned baseplate is unknown.The surgeon revised the baseplate, glenosphere and liner.The surgery was completed successfully.(b)(4).On (b)(6) 2026, the patient came in reporting pain due to a loose baseplate due to poor bone quality.The surgeon revised the baseplate, the glenosphere and the liner.The surgery was completed successfully.Presently, on (b)(6) 2026 the patient came in reporting pain due to a loose baseplate and the cause is unknown.The surgeon revised the reverse hc liner ø42/+6mm to a rev.Constrained e-cross liner ø42/+0, revised the ø42x24.5 glenosphere to a same size glenosphere, and revised the ø24.5x20 - full wedge 10° baseplate to a same size baseplate.The surgery was completed successfully.
|
|
Batch review performed on 2 september 2026.Reverse shoulder system 04.01.0400 grs baseplate - ø24.5x20 - full wedge 10° (k231911) lot 2339042: (b)(4) items manufactured and released on 27-jan-2025.Expiration date: 08-jan-2030.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Root cause:
the event is consistent with aseptic loosening of the glenoid baseplate, which may result from multifactorial clinical and/or mechanical conditions.Humeral impingement against the glenosphere may represent a possible contributing mechanism; however, this cannot be confirmed based on the available information.No device-related root cause or device deficiency has been identified.
|