It was reported that pump battery would not charge even when customer used tandem charging cable, tandem wall adapter, and working wall outlet, and issue did not resolve after leaving pump connected for at least 30 minutes; later customer found damage to charging cable and confirmed pump began charging after switching to different cable and wall adapter.There was no adverse impact to the customer¿s blood glucose level.Customer resolved issue by using non-tandem charging cable with tandem wall adapter, was informed that non-tandem charging supplies were not recommended except for troubleshooting, acknowledged guidance, and was arranged to receive replacement charging supplies, with no pump shutdown and no further assistance required.
|