|
The customer reported that the item is giving false negatives on the bup; according to the facility, "patient is a suboxone patient, tested negative for suboxone, used another cup he had from mckesson and the suboxone was positive." at this time, a sample(s) was not received for investigation; investigation included review of recent complaint trends/scar activity for the last three trend reports, and per vendor, through internal investigation (including dhr review, inspection records review, etc.) no abnormalities were found.The complaint was not confirmed.No further action was taken because without the sample or picture, it is difficult to truly identify the root cause of the issue; the division will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|