• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; TEST,HCG,PREGNANCY,COMBO,CASS,20/10 MIU

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; TEST,HCG,PREGNANCY,COMBO,CASS,20/10 MIU Back to Search Results
Catalog Number MPH22025HCG
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/03/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "we have now come across 2 boxes of medline branded pregnancy tests that have given us false positives on now 4 patients.We had instances where we told the patient they were pregnant and they were not.".
 
Additional Manufacturer Narrative
The customer reported medline pregnancy tests were giving false positives, and stated, "we had instances where we told the patient they were pregnant and they were not"; the customer later stated the tests had given false positives on 5 patients, patients were upset, and no serious injury, medical intervention, or follow-up care was required.A sample was not received for investigation, and retained devices from the reported lot numbers were tested with hcg-negative clinical urine and hcg-negative clinical serum samples; all devices yielded valid negative results and no false positive results were observed during in-house testing.The complaint was not confirmed, and manufacturing batch record review did not uncover any relevant non-conformances and found that the lots met quality control specifications.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retention product, and no indications of a systemic issue were identified.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
TEST,HCG,PREGNANCY,COMBO,CASS,20/10 MIU
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
rahul kota
three lakes drive
northfield, IL 60093--275
MDR Report Key50947015
Report Number1417592-2026-54372
Device Sequence Number9499641
Product Code JHI
UDI-Device Identifier10884389191880
UDI-Public10884389191880
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K062361
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMPH22025HCG
Device Lot Number0000967206
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2025
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-