The customer reported medline pregnancy tests were giving false positives, and stated, "we had instances where we told the patient they were pregnant and they were not"; the customer later stated the tests had given false positives on 5 patients, patients were upset, and no serious injury, medical intervention, or follow-up care was required.A sample was not received for investigation, and retained devices from the reported lot numbers were tested with hcg-negative clinical urine and hcg-negative clinical serum samples; all devices yielded valid negative results and no false positive results were observed during in-house testing.The complaint was not confirmed, and manufacturing batch record review did not uncover any relevant non-conformances and found that the lots met quality control specifications.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retention product, and no indications of a systemic issue were identified.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|