• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; MATTRESS,SYSTEM,A10,LAL/ALTERNATING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; MATTRESS,SYSTEM,A10,LAL/ALTERNATING Back to Search Results
Catalog Number MDT24A10
Medical Device Problem Code Unintended Deflation (4061)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/10/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "air mattress is getting deflated from center.Pump shows normal pressure but as soon as patient lies on the mattress it deflates from the center.".
 
Additional Manufacturer Narrative
The reported product problem/issue was does not hold air, and the complaint stated the air mattress was getting deflated from center, the pump shows normal pressure, and as soon as the patient lies on the mattress it deflates from the center.Medline was not provided the sample and was unable to confirm the complaint; the lot number is from an item that was manufactured in feb 2024, and the manufacturer has a 12 month warranty and would not consider this out of box manufacture defect but normal use wear/tear.The reported problem was unconfirmed, and the cause was documented as sample required to determine cause.There was no reported death or serious injury associated with this event, product was not being used for treatment or diagnosis, and a review of the complaint trending report found no negative trends for this defect or product code and no need for further action.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
MATTRESS,SYSTEM,A10,LAL/ALTERNATING
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
kehkashan khan
three lakes drive
northfield, IL 60093--275
MDR Report Key50947130
Report Number1417592-2026-54397
Device Sequence Number10446271
Product Code FNM
UDI-Device Identifier10193489135671
UDI-Public10193489135671
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDT24A10
Device Lot NumberM05240206
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/10/2025
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-