| Model Number |
7841 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Failure to Capture (1081); Fracture (1260); High impedance (1291); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this right atrial (ra) lead with the cardiac resynchronization therapy defibrillator (crt-d) exhibited high out of range pacing impedances.There was loss of capture (loc) noted and noisy signals.Technical services (ts) reviewed and found this to be indicative of a lead fracture.Programming options were provided.This ra lead remains in service.No adverse patient effects were reported.
|
| |
|
Search Alerts/Recalls
|