The customer reported "many of the protection stations breaking or coming in broken" and that "inr has had 4 manifold break while setting up the table and 1 came in broken in the package." photos were received for review from the customer, a stock check was performed for all available batches, the finished good bill of materials/pack build, manufacturing instructions, drawing, device history records, and receiving inspection records were reviewed.Based on review of the photos, the reported defect was confirmed as the protection station being broken/damaged, and based on the investigation results, the reported event was confirmed as the protection station being damaged/broken at the bond site.Root cause remains inconclusive at this time; no further action is required at this time, and medline will continue to monitor for additional occurrences and adverse trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|