| Model Number |
106524US |
| Medical Device Problem Codes |
Corroded (1131); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bacterial Infection (1735); Purulent Discharge (1812)
|
| Date of Event |
09/18/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient presented with bright blue/green drainage at the driveline exit site (dles), suspectected to be secondary to a confirmed pseudomonas culture which was obtained from the dles.Internal driveline corrosion was ruled out following irrigation and debridement of the exit site for infectious purposes.Antibiotics were administered.No pump data appeared to be abnormal.Upon review of the submitted log files, no data indicated any electrical issues with the driveline.There were no unusual events recorded in the log file event history.The mechanical circulatory support (mcs) equipment was noted to be operating as intended.The event was noted to be on track towards resolution, but not yet resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation conclusion:
the pump remains in use supporting the patient.A correlation between the device and the report of infection could not be conclusively determined.Infection is listed in the instructions for use as a potential adverse event that may be associated with the use of the heartmate 3 left ventricular assist system.Infection has been previously investigated and will continue to be monitored through quality data reviews, which are conducted on production and post product signals to evaluate if products are conforming to product requirements.No further information was provided.The manufacturer is closing the file on this event.
|
| |
|
Search Alerts/Recalls
|