| Model Number |
106524US |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient had frequent low flow alarms that did not appear to be correlated to any activity.The next was to obtain a computed tomography to assess for extrinsic outflow graft obstruction (eogo).Log files were sent for review.The log file captured some intermittent low flow event on 21sep2026.The calculated flow appeared to have fluctuated below the alarm threshold of 2.5 lpm during the events.The log file contained low flow estimates only when the calculated pulsatility index (pi) was dynamic and elevated.
|
| |
|
Additional Manufacturer Narrative
|
|
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
|
| |
|