Model Number 106524US
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/21/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient had frequent low flow alarms that did not appear to be correlated to any activity.The next was to obtain a computed tomography to assess for extrinsic outflow graft obstruction (eogo).Log files were sent for review.The log file captured some intermittent low flow event on 21sep2026.The calculated flow appeared to have fluctuated below the alarm threshold of 2.5 lpm during the events.The log file contained low flow estimates only when the calculated pulsatility index (pi) was dynamic and elevated.
 
Additional Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Thoratec® HeartMate 3® LVAS Implant Kit
Common Device Name
Ventricular (Assist) Bypass
Manufacturer (Section D)
Thoratec Corporation
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
Thoratec Corporation
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
fredric gold
6035 stoneridge drive
pleasanton, CA 94588
+1(781)852-8207
MDR Report Key50947296
Report Number2916596-2026-2920188
Device Sequence Number18754872
Product Code DSQ
UDI-Device Identifier00813024013297
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P160054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/10/2023
Device Model Number106524US
Device Catalogue Number106524US
Device Lot Number8225572
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
Patient Weight286 LB
Patient RaceBlack Or African American
-
-