| Model Number |
G10384 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported, upon insertion of the device during an unknown procedure, a flexor balkin guiding sheath leaked fluid from the valve.A photo provided by the customer shows separation of the hub and separation of the side-arm tubing.Contralateral access was obtained in the femoral artery.The access site was not scarred or calcified, and the anatomy was not stenosed, calcified, or tortuous.The bifurcation was not steep or calcified.No other devices were used through the sheath during the procedure.Resistance was not encountered during insertion/advancement or removal of the sheath.The leak, hub separation, and side-arm separation occurred during the ¿insertion stage¿.The device was not pulled by the side-arm during the procedure.Another device of the same type, from a different lot, was used to successfully complete the procedure.The patient did not require treatment for blood loss.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures or experience any adverse effects due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unknown or unavailable.E1: (b)(6)
h3: the device has been returned, and preliminary evaluation has been performed; however, our investigation is ongoing and device evaluation summary will be included in our follow up report once our investigation has been completed.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Search Alerts/Recalls
|
|
|