• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Nihon Kohden Corporation PU-681RA; Central Monitoring System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Nihon Kohden Corporation PU-681RA; Central Monitoring System Back to Search Results
Model Number PU-681RA
Medical Device Problem Codes No Display/Image (1183); Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/04/2026
Type of Reportable Event Malfunction
Event or Problem Description
The biomedical engineer (bme) reported that the central nurse station (cns) was displaying a "no video present" message on all its monitors.No patient harm was reported.
 
Additional Manufacturer Narrative
The biomedical engineer (bme) reported that the central nurse station (cns) was displaying a "no video present" message on all its monitors.Technical support (ts) had them perform a force shutdown of the cns, rebooted the kvm sender and one of the kvm receivers, but despite repeating the troubleshooting steps multiple times, they were unable to restore video.No patient harm was reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.Additional device information: d10 concomitant medical device: the following device was used in conjunction with the cns: keyboard/video/mouse (kvm) model #:gem-x serial #: ni device manufacturer data: ni unique identifier (udi) #: ni returned to nihon kohden: na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PU-681RA
Common Device Name
Central Monitoring System
Manufacturer (Section D)
Nihon Kohden Corporation
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
Nihon Kohden Tomioka Corporation
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key50947454
Report Number8030229-2026-07197
Device Sequence Number9624036
Product Code MHX
UDI-Device Identifier04931921131640
UDI-Public(01)04931921131640
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K102376
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPU-681RA
Device Catalogue NumberPU-681RA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/02/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
KVM
Patient SexUnknown
-
-