• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN? X4 CRT-D; Defibrillator, Automatic Implantable Cardioverter, With Crt-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN? X4 CRT-D; Defibrillator, Automatic Implantable Cardioverter, With Crt-D Back to Search Results
Model Number G148
Medical Device Problem Codes Pacing Problem (1439); Defibrillation/Stimulation Problem (1573); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Codes Tachycardia (2095); Shock from Patient Lead(s) (3162)
Date of Event 09/08/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) delivered potentially inappropriate anti-tachycardia pacing (atp) and a shock.The onset of the arrythmia appeared to possibly be atrial flutter with rapid ventricular response (rvr).The atp possibly put the patient into ventricular tachycardia (vt).The shock delivered converted the vt, but the atrial flutter remained.Technical services (ts) discussed possible reprogramming options for this device.This crt-d remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN? X4 CRT-D
Common Device Name
Defibrillator, Automatic Implantable Cardioverter, With Crt-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50947458
Report Number2124215-2026-51994
Device Sequence Number18547679
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public(01)00802526534591(17)220207
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/07/2022
Device Model NumberG148
Device Catalogue NumberG148
Device Lot Number234131
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2026
Initial Report FDA Received Date09/30/2026
Date Device Manufactured02/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age80 YR
Patient SexMale
-
-