| Model Number |
G148 |
| Medical Device Problem Codes |
Pacing Problem (1439); Defibrillation/Stimulation Problem (1573); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Codes |
Tachycardia (2095); Shock from Patient Lead(s) (3162)
|
| Date of Event |
09/08/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) delivered potentially inappropriate anti-tachycardia pacing (atp) and a shock.The onset of the arrythmia appeared to possibly be atrial flutter with rapid ventricular response (rvr).The atp possibly put the patient into ventricular tachycardia (vt).The shock delivered converted the vt, but the atrial flutter remained.Technical services (ts) discussed possible reprogramming options for this device.This crt-d remains in service.No adverse patient effects were reported.
|
| |
|
Search Alerts/Recalls
|