| Model Number |
PT-000409 |
| Medical Device Problem Codes |
Overheating of Device (1437); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient¿s omnipod 5 controller/personal diabetes manager (pdm) becomes warm to the touch when charging and loses charge faster than expected.It was reported that the heated area appears to be where the battery is located in the device.No harm to the patient was reported and no medical assistance was required.A replacement controller and charging cable have been issued to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
According to the complainant, the device will not be returned for investigation.We are unable to confirm the cause of the reported event.The reported thermal device malfunction is associated with a personal diabetes manager manufactured after the correction for action res#91127 was implemented.Lot release records were reviewed and the product lot met all acceptance criteria.Locked down smartphone: lockdown
omnipod software app version: 4.0.3
operating system: n5004l-am-q-mv01602-06-01.06
hardware: n5004l
cgm sensor type: data not available
please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
|
| |
|