• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Roche Diagnostics Triglycerides; Triglyceride test system

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Roche Diagnostics Triglycerides; Triglyceride test system Back to Search Results
Catalog Number 20767107322
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/08/2026
Type of Reportable Event Malfunction
Event or Problem Description
We received an allegation about discrepant results for 1 patient sample tested with triglycerides assay on a cobas 6000 c (501) module.The initial result was 594 mg/dl.The customer noticed that the initial result was high and did not match the patient's clinical picture or previous results, prompting a repeat of the sample.The repeat results were: 788 mg/dl (accompanied by an invalidating data flag).5651 mg/dl (accompanied by an invalidating data flag).4425 mg/dl.The repeat result of 4425 mg/dl was deemed to be correct.
 
Additional Manufacturer Narrative
The triglycerides reagent expiration date was not provided.The cobas 6000 c 501 module serial number was (b)(6).The visual inspection of the sample showed that it was lipemic.The investigation is ongoing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Triglycerides
Common Device Name
Triglyceride test system
Manufacturer (Section D)
Roche Diagnostics
9115 hague road
indianapolis IN 46250 -045
Manufacturer (Section G)
Roche Diagnostics GmbH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
natalie catton
9115 hague road
indianapolis, IN 46250
MDR Report Key50947532
Report Number1823260-2026-04026
Device Sequence Number9499660
Product Code CDT
UDI-Device Identifier04015630918393
UDI-Public04015630918393
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K873049
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number20767107322
Device Lot Number951776
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-