| Catalog Number |
20767107322 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
We received an allegation about discrepant results for 1 patient sample tested with triglycerides assay on a cobas 6000 c (501) module.The initial result was 594 mg/dl.The customer noticed that the initial result was high and did not match the patient's clinical picture or previous results, prompting a repeat of the sample.The repeat results were:
788 mg/dl (accompanied by an invalidating data flag).5651 mg/dl (accompanied by an invalidating data flag).4425 mg/dl.The repeat result of 4425 mg/dl was deemed to be correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The triglycerides reagent expiration date was not provided.The cobas 6000 c 501 module serial number was (b)(6).The visual inspection of the sample showed that it was lipemic.The investigation is ongoing.
|
| |
|
Search Alerts/Recalls
|