(b)(4)
date sent: 9/30/2026
d4: batch #unk.Additional information was requested, and the following was obtained:
was the device locked on tissue with the jaws closed? if yes, what steps were taken to open the jaws? (pull open the closing trigger, press home button, use bailout) all the above
if the jaws could not be opened, how was the device removed manual override
did the jaws eventually open? no could you please clarify if there were any patient consequences? surgeon corrected the previous report, there was no patient harm
could you please clarify if there were any change in the post-operative care of the patient as a result of the event? there was not
could you please provide more details regarding how the device was removed from the patient tissue using the manual override? device fired, successfully deploying the staples and cutting the tissue.The device wouldn't open, it was unknown if the blade retracted safely, so manual override was ratcheted to be certain.Was the device successfully removed without opening the jaws, or was the tissue cut in order to remove the device? successfully removed
you indicated that the device was removed using the manual override; however, you also stated that the jaws did not eventually open.Could you please clarify whether the jaws opened at any point during the removal process? if so, please describe the actions or steps taken the jaws did not open.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device psee45a lot number a9m27t and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|