Catalog Number MX6721
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
As reported, the polyethylene glycol (peg) plug of a 6/7f mynx grip vascular closure device (vcd) would not deploy after several attempts.Hemostasis was achieved by manual compression.There was no reported patient injury.The user shuttled down completely, and there was no significant resistance when shuttling down.The advancer tube was not engaged or visible.No damage was found on the device or device packaging prior to opening.The device was stored and prepared in accordance with the instructions for use (ifu).A retrograde approach was used.The deployer was mynx certified.The vessel type was arterial, the vessel diameter was verified to be greater than or equal to 5 mm, and the stick location was the right femoral artery.The device will not be returned for evaluation as it was discarded.
 
Additional Manufacturer Narrative
Due to system limitations, section h6 required to input type of investigation, investigation findings, and investigation conclusions for an initial report.Pending additional information to complete investigation.Additional information is pending and will be submitted within 30 days upon receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYNXGRIP
Common Device Name
Device, Hemostasis, Vascular
Manufacturer (Section D)
Cordis US. Corp
14201 nw 60th avenue
miami lakes FL 33014
Manufacturer (Section G)
Cordis US Corp.
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
5452 betsy ross drive
santa clara, CA 95054
+1(786)313-8372
MDR Report Key50947596
Report Number3004939290-2026-01363
Device Sequence Number10348732
Product Code MGB
UDI-Device Identifier10862028000410
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P040044
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMX6721
Device Lot NumberF2523302
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Mynx Arterial
Patient SexUnknown
-
-