• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) Assert-IQ¿ ICM; Recorder, Event, Implantable Cardiac, (Without Arrhythmia Detection)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) Assert-IQ¿ ICM; Recorder, Event, Implantable Cardiac, (Without Arrhythmia Detection) Back to Search Results
Model Number DM5500
Medical Device Problem Codes Premature Discharge of Battery (1057); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/19/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient's insertable cardiac monitor (icm) was suspected to exhibit premature battery depletion due to an incorrect measurement of battery longevity that later returned to its appropriate value.Afterwards, it was noted that icm exhibited premature elective replacement indicator (eri) and was confirmed to exhibit early battery depletion.The icm was explanted and replaced.The patient had no consequences due to the event.
 
Additional Manufacturer Narrative
The reported complaint of premature battery depletion was confirmed.Based on the information provided, it was determined there was unexpected increase in the current drain and drop in the battery voltage, resulting in battery depletion.The root cause of the unexpected increase in current drain and drop in voltage could not be determined with the available data.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Assert-IQ¿ ICM
Common Device Name
Recorder, Event, Implantable Cardiac, (Without Arrhythmia Detection)
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50947602
Report Number2017865-2026-20715
Device Sequence Number13349385
Product Code MXC
UDI-Device Identifier05415067047809
UDI-Public(01)05415067047809(10)S000092892(17)250608
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K230286
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/08/2025
Device Model NumberDM5500
Device Lot NumberS000092892
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2026
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexMale
-
-