| Catalog Number |
00515048200 |
| Medical Device Problem Code |
Inability to Irrigate (1337)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during surgery brand new device did not spray water.Investigation results found the batteries had leaked electrolyte inside of the pack.There was no patient impact.Due diligence is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).G2: event occurred in china.Visual examination of the returned product identified evidence of electrolyte leakage inside the battery pack with no apparent damage to pack wiring; the batteries were received removed from the pack.Review of the device history record(s) identified no deviations or anomalies during manufacturing.The root cause is attributed to a manufacturing issue exacerbated by a design issue.Able to confirm reported event.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|