It was reported that patient experienced bulging and disfigurement in the penile body with the inflatable penile prosthesis (ipp).Upon revision, physician noticed that the reservoir wasn't placed correctly and caused a bulge in the abdominal cavity.Per additional information, it was later reported that also the polyester braided layer inside the cylinder ruptured, resulting in column deformation.As a result, the device was explanted and replaced.No further complications were reported.
|
|
The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Based on the information available, no problem was detected that contributed to the reported event.Therefore, a conclusion code of no problem detected was assigned to this investigation.Model number/catalog number: 72404014,
serial number: n/a,
batch/lot number: 1100348585,
model/catalog description: cxr 18cm,
unique identifier (udi)#: (b)(4).
|