According to the customer, they found that it was difficult to return the stylets into several 18g spinal needles and that it "felt like rubbing sandpaper together"; the customer reported that no patients were "affected by the spinal needle" and there was no serious injury, medical intervention, or follow-up care.Medline quality received samples, visually inspected them for damage, defects and deformities, and evaluated the stylet needles by inserting and reinserting the stylet needle into the sheath.No defects were found during visual inspection; some stylet needles inserted smoothly, while others exhibited scraping during insertion, and medline was able to confirm the complaint.Based on the sample and complaint details, the root cause is considered to be a vendor product issue relating to the manufacturing process; the quality team will monitor this issue for trending purposes and the complaint has been provided to the vendor/manufacturer for further evaluation.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|