• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; METER,GLUCOSE,EVENCARE G3,PROF.USE ONLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; METER,GLUCOSE,EVENCARE G3,PROF.USE ONLY Back to Search Results
Catalog Number MPH3540
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that unit has failed api tesing.They use american proficiency institute labs to test the monitors quarterly, and this unit failed their testing.
 
Additional Manufacturer Narrative
The customer reported the unit failed american proficiency institute testing, and customer does test weekly using medline's control solution, and it is within range using the control solution.Medline received 1 ea evencare g3 blood glucose meter in used condition, inserted a control cr2032 battery, and proceeded to test the sample according to the instructions for use using test strips and control solutions.Results were within control solutions range, there were no indications power or display issues, and the complaint was not confirmed.The sample turned on and displayed readings within the test settings, the sample worked as intended, and the division will continue to monitor this issue for trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
METER,GLUCOSE,EVENCARE G3,PROF.USE ONLY
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
nathan bersch
three lakes drive
northfield, IL 60093--275
MDR Report Key50947854
Report Number1417592-2026-54504
Device Sequence Number10348739
Product Code NBW
UDI-Device Identifier10884389831199
UDI-Public10884389831199
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K120448
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMPH3540
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer05/22/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2024
Initial Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-